Q

How many measurement sampling frequencies does the waveform measurement function support?

A

The measurement sampling frequency and time vary with the instrument being used. The instrument determines the optimal measurement sampling frequency to use based on the waveform being measured.

For example, when measuring a 60 Hz sine wave with th CM4372, the sampling frequency will be 2.88 kHz, and the sampling time will be 50 ms.

Since the period of a 60 Hz sine wave is 16.666 ms, 3 wavelengths will be displayed.

Since the CM4371-50 and CM4373-50 have a measurement time of 50 ms, you can display 3 cycles of a 60 Hz waveform.

Although this information is not part of the product specifications, instruments can measure the waveforms such as the following:

CM4371-50, CM4373-50, CM4375-50, CM4141-50, CM4372, CM4374, CM4376, CM4142, DT4261


AC/DC CLAMP METER CM4371-50
Measurement signal frequency Sampling frequency Sampling time
0 Hz ≦ freq < 40 Hz 1.44 kHz 100 ms
40 Hz ≤ freq < 80 Hz 2.88 kHz 50 ms
80 Hz ≤ freq < 160 Hz 5.76 kHz 25 ms
160 Hz ≤ freq < 390 Hz 14.4 kHz 10 ms
390 Hz ≤ freq < 1000 Hz 28.8 kHz 5 ms

CM3286-50, CM3286-01


Measurement signal frequency Sampling frequency Sampling time
freq ≦ 55 Hz 12.8 kHz 20 ms
freq > 55 Hz 15.36 kHz 16.6 ms

CM4001


Measurement signal frequency Sampling frequency Sampling time
40 Hz ≤ freq < approx. 80 Hz 2.88 kHz 50 ms
Approx. 80 Hz ≤ freq < approx. 160 Hz 5.76 kHz 25 ms
Approx. 160 Hz ≤ freq < approx. 390 Hz 14.4 kHz 10 ms
Approx. 390 Hz ≤ freq < 1000 Hz 28.8 kHz 5 ms

CM4002, CM4003


Measurement signal frequency Sampling frequency Sampling time
40 Hz ≤ freq < approx. 80 Hz 2.4 kHz 50 ms
Approx. 80 Hz ≤ freq < approx. 160 Hz 4.8 kHz 25 ms
Approx. 160 Hz ≤ freq < approx. 390 Hz 12 kHz 10 ms
Approx. 390 Hz ≤ freq < 1000 Hz 24 kHz 5 ms

FT6380-50


Measurement signal frequency Sampling frequency Sampling time
30 Hz ≤ freq < 40 Hz 1 kHz 160 ms
40 Hz ≤ freq < 80 Hz 2 kHz 80 ms
80 Hz ≤ freq < 160 Hz 4 kHz 40 ms
160 Hz ≤ freq < 400 Hz 4 kHz 20 ms

CM7291


Measurement signal frequency Sampling frequency Sampling time
0 Hz ≤ freq < approx. 40 Hz 500 Hz 200 ms
Approx. 40 Hz ≤ freq < approx. 80 Hz 1 kHz 100 ms
freq ≥ approx. 80 Hz 2 kHz 50 ms