Q

What’s the shortest event duration that can be detected by the event recording function?

A

Specifications vary by instrument.

Which measurement functions can be used by the event recording function depends on the instrument.

In addition, threshold value settings, the maximum number of events that can be saved, and the minimum detectable event duration vary from instrument to instrument.

Please exercise care concerning each instrument’s specifications.

  • When the maximum number of events that can be saved is recorded, measurement will stop, and subsequent events will not be recorded.
  • It may not be possible to detect events that last less than the minimum detectable event duration due to the inability to measure them accurately.

CM4001


AC LEAKAGE CLAMP METER CM4001
  • Threshold value settings: > (Greater than) only
  • Maximum number of events that can be saved: 99
Supported functions Minimum detectable event duration
AC current measurement (ACA) 200ms

CM4002, CM4003


  • Threshold value settings: > (Greater than), < (Less than)
  • Maximum number of events that can be saved: 999
Supported functions Minimum detectable event duration
AC current measurement (ACA) 200ms

FT6380-50


  • Threshold value settings: > (Greater than), < (Less than)
  • Maximum number of events that can be saved: 25
Supported functions Minimum detectable event duration
AC current measurement (ACA) 1s
Resistance measurement (Ω)

DT4261


  • Threshold value settings: > (Greater than), < (Less than)
  • Maximum number of events that can be saved: 99
Supported functions Minimum detectable event duration
DC voltage measurement (DCV) 200ms
DC voltage measurement (HIGH V PROBE mode) (DCHV)
AC voltage measurement (ACV)
AC + DC voltage measurement (ACDCV)
Voltage frequency measurement (HzV)
AC current measurement (clamp sensor use) (Clamp)
AC current measurement (ACA)
DC current measurement (DCA)
AC + DC current measurement (ACDCA)
Current frequency measurement (HzA)
Resistance measurement (Ω)
Capacitance measurement (F)

CM4371-50, CM4373-50, CM4375-50


  • Threshold value settings: > (Greater than), < (Less than)
  • Maximum number of events that can be saved: 99
Supported functions Minimum detectable event duration
DC voltage measurement (DCV) 400ms
DC voltage measurement (HIGH V PROBE mode) (DCHV)
AC voltage measurement (ACV)
AC + DC voltage measurement (ACDCV)
Voltage frequency measurement (HzV) 400 ms to 4000 ms (varies with measured value)
AC current measurement (ACA) 400ms
DC current measurement (DCA)
AC + DC current measurement (ACDCA)
Current frequency measurement (HzA) 400 ms to 4000 ms (varies with measured value)
Resistance measurement (Ω) 400ms
Capacitance measurement (F) 400 ms to 4000 ms (varies with measured value)
Temperature measurement (°C/°F) 2000ms

CM4141-50


  • Threshold value settings: > (Greater than), < (Less than)
  • Maximum number of events that can be saved: 99
Supported functions Minimum detectable event duration
DC voltage measurement (DCV) 400ms
DC voltage measurement (HIGH V PROBE mode) (DCHV)
AC voltage measurement (ACV)
AC + DC voltage measurement (ACDCV)
Voltage frequency measurement (HzV) 400 ms to 4000 ms (varies with measured value)
AC current measurement (ACA) 400ms
Current frequency measurement (HzA) 400 ms to 4000 ms (varies with measured value)
Resistance measurement (Ω) 400ms
Capacitance measurement (F) 400 ms to 4000 ms (varies with measured value)
Temperature measurement (°C/°F) 2000ms

CM3286-50


  • Threshold value settings: > (Greater than), < (Less than)
  • Maximum number of events that can be saved: 99
Supported functions Minimum detectable event duration
AC current measurement (ACA) 1s
AC voltage measurement (ACV)