Q
What’s the shortest event duration that can be detected by the event recording function?
A
Specifications vary by instrument.
Which measurement functions can be used by the event recording function depends on the instrument.
In addition, threshold value settings, the maximum number of events that can be saved, and the minimum detectable event duration vary from instrument to instrument.
Please exercise care concerning each instrument’s specifications.
- When the maximum number of events that can be saved is recorded, measurement will stop, and subsequent events will not be recorded.
- It may not be possible to detect events that last less than the minimum detectable event duration due to the inability to measure them accurately.
CM4001
- Threshold value settings: > (Greater than) only
- Maximum number of events that can be saved: 99
Supported functions | Minimum detectable event duration |
---|---|
AC current measurement (ACA) | 200ms |
CM4002, CM4003
- Threshold value settings: > (Greater than), < (Less than)
- Maximum number of events that can be saved: 999
Supported functions | Minimum detectable event duration |
---|---|
AC current measurement (ACA) | 200ms |
FT6380-50
- Threshold value settings: > (Greater than), < (Less than)
- Maximum number of events that can be saved: 25
Supported functions | Minimum detectable event duration |
---|---|
AC current measurement (ACA) | 1s |
Resistance measurement (Ω) |
DT4261
- Threshold value settings: > (Greater than), < (Less than)
- Maximum number of events that can be saved: 99
Supported functions | Minimum detectable event duration |
---|---|
DC voltage measurement (DCV) | 200ms |
DC voltage measurement (HIGH V PROBE mode) (DCHV) | |
AC voltage measurement (ACV) | |
AC + DC voltage measurement (ACDCV) | |
Voltage frequency measurement (HzV) | |
AC current measurement (clamp sensor use) (Clamp) | |
AC current measurement (ACA) | |
DC current measurement (DCA) | |
AC + DC current measurement (ACDCA) | |
Current frequency measurement (HzA) | |
Resistance measurement (Ω) | |
Capacitance measurement (F) |
CM4371-50, CM4373-50, CM4375-50
- Threshold value settings: > (Greater than), < (Less than)
- Maximum number of events that can be saved: 99
Supported functions | Minimum detectable event duration |
---|---|
DC voltage measurement (DCV) | 400ms |
DC voltage measurement (HIGH V PROBE mode) (DCHV) | |
AC voltage measurement (ACV) | |
AC + DC voltage measurement (ACDCV) | |
Voltage frequency measurement (HzV) | 400 ms to 4000 ms (varies with measured value) |
AC current measurement (ACA) | 400ms |
DC current measurement (DCA) | |
AC + DC current measurement (ACDCA) | |
Current frequency measurement (HzA) | 400 ms to 4000 ms (varies with measured value) |
Resistance measurement (Ω) | 400ms |
Capacitance measurement (F) | 400 ms to 4000 ms (varies with measured value) |
Temperature measurement (°C/°F) | 2000ms |
CM4141-50
- Threshold value settings: > (Greater than), < (Less than)
- Maximum number of events that can be saved: 99
Supported functions | Minimum detectable event duration |
---|---|
DC voltage measurement (DCV) | 400ms |
DC voltage measurement (HIGH V PROBE mode) (DCHV) | |
AC voltage measurement (ACV) | |
AC + DC voltage measurement (ACDCV) | |
Voltage frequency measurement (HzV) | 400 ms to 4000 ms (varies with measured value) |
AC current measurement (ACA) | 400ms |
Current frequency measurement (HzA) | 400 ms to 4000 ms (varies with measured value) |
Resistance measurement (Ω) | 400ms |
Capacitance measurement (F) | 400 ms to 4000 ms (varies with measured value) |
Temperature measurement (°C/°F) | 2000ms |
CM3286-50
- Threshold value settings: > (Greater than), < (Less than)
- Maximum number of events that can be saved: 99
Supported functions | Minimum detectable event duration |
---|---|
AC current measurement (ACA) | 1s |
AC voltage measurement (ACV) |